Skip to main content

Table 4 Summary of average enzymes per bead for prototype assay calibration using Fujirebio standard for assay evaluation and clinical test batches

From: A digital enzyme-linked immunosorbent assay for ultrasensitive measurement of amyloid-β 1–42 peptide in human plasma with utility for studies of Alzheimer’s disease therapeutics

 

AEB

Theoretical concentration, pg/ml

0.023

0.069

0.206

0.617

1.85

5.6

16.7

50.0

 Run 6

0.006

0.007

0.008

0.012

0.031

0.164

1.105

8.401

 Run 7

0.011

0.012

0.014

0.015

0.036

0.142

1.083

8.115

 Run 8

0.011

0.012

0.013

0.013

0.033

0.149

0.931

8.084

N (total number of runs)

3

3

3

3

3

3

3

3

Inter-assay mean

0.009

0.010

0.012

0.013

0.033

0.151

1.040

8.200

SD

0.003

0.003

0.003

0.002

0.002

0.011

0.095

0.175

CV%

28.8

29.1

25.0

13.8

6.6

7.3

9.1

2.1

  1. AEB Average enzyme per bead, CV% Coefficient of variation percentage
  2. In runs 6–8, calibration curves prepared using both Quanterix and Fujirebio standards were included