Table 2 Prototype assay calibration performance summary using Fujirebio standard for assay evaluation and clinical test batches

From: A digital enzyme-linked immunosorbent assay for ultrasensitive measurement of amyloid-β 1–42 peptide in human plasma with utility for studies of Alzheimer’s disease therapeutics

 

Concentrationa (pg/ml)

Theoretical concentration

0.023

0.069

0.206

0.617

1.85

5.6

16.7

50.0

 Run 6

−0.006

0.097

0.262

0.584

1.83

5.65

16.6

50.1

 Run 7

 

0.112

0.365

0.514

1.94

5.36

17.1

49.8

 Run 8

 

0.260

0.283

0.388

1.98

5.79

16.0

50.5

N (total number of runs)

1

3

3

3

3

3

3

3

Inter-assay mean

−0.006

0.156

0.303

0.495

1.92

5.6

16.6

50.1

SD

 

0.090

0.054

0.100

0.077

0.220

0.508

0.365

CV%

 

57.9

17.9

20.1

4.0

3.9

3.1

0.7

Inter-assay RE, %

−125

128

47.5

−19.8

3.6

0.8

−0.7

0.3

  1. CV% Coefficient of variation percentage, RE Relative error
  2. In runs 6–8, calibration curves prepared using both Quanterix and Fujirebio standards were included
  3. aConcentrations were interpolated using the calibrator average enzymes per bead values against the fitted cubic curves